TEST ID HITPANEL HIT (HEPARIN-INDUCED THROMBOCYTOPENIA) PF4 PANEL

Important Note

If down:

OU- HIT Heparin- Induced Platelet Antibody

RML- Test number- 5566580

CPL- Test number-3732

Additional Codes

PF4

Performing Laboratory

NRLS-Core Coagulation

Specimen Type

Citrate Plasma AND Serum

Specimen Required

*Container/Tube:
Preferred:  3.2% blue top sodium citrate tube, 1.8 or 2.7 mL
Specimen Volume: Tube must be full

 

*Container/Tube:
Preferred:  Red Top No gel 
Specimen Volume: 5 ml

Collection Instructions:

Plasma gel tubes should be centrifuged within 2 hours of collection. 4 hours is acceptable.
Red-top tubes should be fully clotted, centrifuged and aliquoted within 2 hours of collection. 4 hours is acceptable.

 

Patients with high hematocrit (greater than 55%) results will need a citrate adjusted blue top tube from the laboratory for this test. 

Specimen Minimum Volume

Blue top-Tube must be full

Red top- 1.0 mL

Specimen Stability Information

Specimen Type

Temperature

Time 

Citrated Plasma or Whole Blood Refrigerated(preferred) 4 hours
Citrated Plasma or Whole Blood Ambient 4 hours
Citrated Plasma Frozen 2 months
     
Serum(red top only) Refrigerated(preferred) 1 week 
  Frozen Indefinite

Profile Information

Reporting Name

Always Performed

Anti-Platelet Factor 4-Heparin (PF4/H) Ab Yes

Reflex Testing

Reporting Name

Available Separately

Always Performed

SRA (Serotonin Release Assay) Yes No

*A positive HIT-Ab result will reflex a SRA (Serotonin Release Assay) to be sent to reference laboratory for the confirmation or exclusion of HIT Ab.

Reference Values

An interpretive report will be provided